Schiller 10 leads ECG/EKG cable with medical snaps

Production Feature

1:Insulated performance:Between wire and wire ,wire and shield ,the insulated resistance should >100MΩ ,the testing voltage should be 100 DC.

2:Strength of anti electricity , between wire and wire ,wire and shield,exert AC1000V±20%,50Hz voltage,1 s, creepage current<0.25mA, 1 s, creepage current≤0.25mA

3:Creepage current: use the 120V AC 60HZ power supply, live wire tadem ammeter and all trunk cable and lead wire signal cable, the ammeter indicate the passed creepage current less than 10A

4:Anti defibrillation performance(sutible for the product with defibrillation function) The product suffer DC5000V 360J(joule)20% power surge 3 times (every time exert electricity power time is 60seconds) Change in resistance value does not exceed ±10%